Examination of needle surface corrosion in electroacupuncture
  1. Claire Shuiqing Zhang1,
  2. George Shengxi Zhang1,
  3. Shanqing Xu2,
  4. Bin Li3,
  5. Anthony Lin Zhang1,
  6. Charlie Changli Xue4,
  7. Yi Min Xie2
  1. 1Discipline of Chinese Medicine, School of Health and Biomedical Sciences, RMIT University, VIC 3083, Australia
  2. 2Centre for Innovative Structures and Materials, School of Engineering, RMIT University, VIC 3083, Australia
  3. 3Department of Acupuncture, Beijing Hospital of Chinese Medicine, Beijing, China
  4. 4School of Health and Biomedical Sciences, RMIT University, VIC 3083, Australia
  1. Correspondence to Dr Claire Shuiqing Zhang, Discipline of Chinese Medicine, School of Health and Biomedical Sciences RMIT University VIC 3083 Australia; claire.zhang{at}rmit.edu.au

Abstract

Background Electroacupuncture (EA) is a modern form of acupuncture therapy where stainless steel acupuncture needles are used as percutaneous electrodes to apply electrical stimulation. The concern about electrolytic corrosion of needles during EA has not been conclusively addressed.

Aim To examine whether corrosion of stainless steel acupuncture needles occurs after EA to establish the safety profile of this therapy.

Methods The study comprised four EA sessions on healthy participants mimicking a common clinical practice, with needle surface examinations conducted immediately after each session. Used acupuncture needles that did not undergo electrical stimulation and unused needles taken from the original package were also examined as control comparisons. Two commonly used types of single-use, silicone-coated, stainless steel needles were selected for the experiment. The ES-160 digital acupunctoscope (a charge-balanced electric stimulator) was used to deliver electrical stimulation, and an oscilloscope was used to record the waveforms and electric currents. All needles were sterilised before examination by a scanning electron microscope (SEM), where images of needle tips and shafts were taken for further analysis.

Results and conclusions 32 needles were examined under the SEM. As the main findings, the SEM images showed the surface finish, burr attachments and surface characteristics of needle samples. No evidence of electrolytic corrosion was detected on any needle that underwent electrical stimulation for 30 min delivered by a charge-balanced acupunctoscope in healthy participants. The choice of a charge-balanced acupunctoscope is recommended to avoid any potential corrosion of needles in EA clinical practice.

  • electroacupuncture
  • adverse events

Statistics from Altmetric.com

Footnotes

  • Contributors CSZ and YMX initiated and designed the research. CSZ and GSZ conducted the clinical treatments. SQX conducted the laboratory needle examination. CSZ, GSZ, SQX and YMX analysed data and drafted the manuscript. BL, ALZ and CCX provided critical comments and contributed to the revision of the draft. All authors read and approved the final version of the manuscript accepted for publication.

  • Funding This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.

  • Competing interests None declared.

  • Ethics approval This project was approved by the RMIT University Human Research Ethics Committee prior to its commencement.

  • Provenance and peer review Not commissioned; externally peer reviewed.